美國 Thermo EUTECH 線上導電度控制器
-
in 美國
此為美國 Thermo EUTECH 線上EC控制器,為中品質 中價位,有各種功能和機種,此EUTECH被THERMO併購 主機為綠色版本。
Conductivity | ||
Range | ... to 1.999 µS/cm ; ... to 19.99 µS/cm ; ... to 199.9 µS/cm ; ... to 1999 µS/cm ; ... to 5000 µS/cm ; ... to 19.99 mS/cm ; ... to 199.9 mS/cm |
|
Resolution | 0.001 µS/cm ; 0.01 µS/cm ; 0.1 µS/cm ; 1 µS/cm ; 5 µS/cm ; 0.01 mS/cm ; 0.1 mS/cm |
|
Accuracy | ±1 % full scale reading | |
Cell Constant | 0.01 ; 0.1 ; 1 ; 10 | |
Temperature | ||
Range | -9.9 to 125 ºC | |
Resolution | 0.1 ºC | |
Accuracy | ±0.5 ºC | |
Sensor | Pt100 / Pt1000 (jumper selectable) | |
Compensation | Auto/manual (normalized at 25 ºC) | |
Coefficient | Ultrapure water or linear 0.00 to 10.00 % | |
Set Point and Controller Functions | ||
Set Point 1 (SP1) / Set Point 2 (SP2) |
... to 1.999 µS/cm or ... to 19.99 µS/cm or ... to 199.9 µS/cm or ... to 5000 µS/cm or ... to 19.99 mS/cm or ... to 199.9 mS/cm |
|
Switching Conductivity Hysteresis | 0 to 10 % of full scale | |
Function (Switchable) | P control (pulse length/pulse frequency); limit control | |
Adjustable Period With Pulse Length Controller |
0.5 to 20 sec | |
Adjustable Period With Pulse Frequency Controller | 60 to 120 pulse/min | |
Pickup/Dropout Delay | 0 to 2000 sec | |
Contact Outputs | 3 SPDT relays | |
Switching Voltage/ Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA | |
Alarm Functions | ||
Function (Switchable) | Steady or fleet (pulse) | |
Wash Cycle |
0.1 to 199.9 hr | |
Wash Duration | 1 to 1999 sec | |
Pickup Delay | 0 to 2000 sec | |
Switching Voltage/ Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA | |
Electrical Data and Connections | ||
Transmitter function | 0/4 to 20 mA scalable outputs for Conductivity, galvanically isolated |
|
Hold function switch | To freeze output current and deactivate control relays | |
Load | Max. 600 Ohm | |
Conductivity input | 2-pin terminal | |
Connection terminal | 5-pole, 17-pole terminal, detachable blocks | |
Power Supply | ||
Input |
110 VAC |
220 VAC (jumper selectable) ; 48 to 62 Hz ; max. 7 VA (For TSCONCTP1002) |
Main Fuse | Slow-blow 250 V / 100 mA | |
Pollution Degree | 2 | |
Transient Overvoltage Category | II | |
EMC Specifications | ||
Emitted Interference | According to EN 50081-1 | |
Immunity to Interference | According to EN 50082-1 | |
Environmental Conditions | ||
Operating temperature range | -10 to 50 ºC | |
Max. relative humidity | 80 % up to 31 ºC decreasing linearly to 50 % at 40 ºC | |
Mechanical Specifications | ||
Dimensions (WxHxD) | 96 x 96 x 175 mm | |
Weight | 700 g (unit) / 800 g (packed) | |
Ingress protection | IP54 (front panel) |